Gamma-ray spectroscopy is the quantitative study of the energy spectra of gamma-ray sources, in such as the nuclear industry, geochemical investigation, and astrophysics.
Most radioactive sources produce gamma rays, which are of various
energies and intensities. When these emissions are detected and analyzed
with a spectroscopy system, a gamma-ray energy spectrum can be
produced.
A detailed analysis of this spectrum is typically used to determine
the identity and quantity of gamma emitters present in a gamma source,
and is a vital tool in radiometric assay. The gamma spectrum is
characteristic of the gamma-emitting nuclides contained in the source, just as in optical spectroscopy, the optical spectrum is characteristic of the material contained in a sample.
System components and principle of operation
The equipment used in gamma spectroscopy includes an energy-sensitive
radiation detector, electronics to process detector signals produced by
the detector, such as a pulse sorter (i.e., multichannel analyzer), and
associated amplifiers and data readout
devices to generate, display, and store the spectrum. Other components,
such as rate meters and peak position stabilizers, may also be
included.
The most common detectors include sodium iodide (NaI) scintillation counters and high-purity germanium detectors.
Gamma spectroscopy detectors are passive materials that wait for a
gamma interaction to occur in the detector volume. The most important
interaction mechanisms are the photoelectric effect, the Compton effect, and pair production.
The photoelectric effect is preferred, as it absorbs all of the energy
of the incident gamma ray. Full energy absorption is also possible when a
series of these interaction mechanisms take place within the detector
volume. When a gamma ray undergoes a Compton interaction or pair
production, and a portion of the energy escapes from the detector volume
without being absorbed, the background rate in the spectrum is
increased by one count. This count will appear in a channel below the
channel that corresponds to the full energy of the gamma ray. Larger
detector volumes reduce this effect.
The voltage pulse produced by the detector (or by the photomultiplier
in a scintillation counter) is shaped by a multichannel analyzer (MCA).
The multichannel analyzer takes the very small voltage signal produced
by the detector, reshapes it into a Gaussian or trapezoidal shape, and converts that signal into a digital signal. In some systems, the analog-to-digital conversion
is performed before the peak is reshaped. The analog-to-digital
converter (ADC) also sorts the pulses by their height. ADCs have
specific numbers of "bins" into which the pulses can be sorted; these
bins represent the channels in the spectrum. The number of
channels can be changed in most modern gamma spectroscopy systems by
modifying software or hardware settings. The number of channels is
typically a power of two; common values include 512, 1024, 2048, 4096,
8192, or 16384 channels. The choice of number of channels depends on the
resolution of the system and the energy range being studied.
The multichannel analyzer output is sent to a computer, which stores,
displays, and analyzes the data. A variety of software packages are
available from several manufacturers, and generally include spectrum
analysis tools such as energy calibration, peak area and net area
calculation, and resolution calculation.
Laboratory equipment for determination of γ-radiation spectrum with a
scintillation counter. The output from the scintillation counter goes to
a Multichannel Analyser which processes and formats the data.
Detector performance
Gamma spectroscopy systems are selected to take advantage of several performance characteristics. Two of the most important include detector resolution and detector efficiency.
Detector resolution
Gamma rays detected in a spectroscopic system produce peaks in the spectrum. These peaks can also be called lines
by analogy to optical spectroscopy. The width of the peaks is
determined by the resolution of the detector, a very important
characteristic of gamma spectroscopic detectors, and high resolution
enables the spectroscopist to separate two gamma lines that are close to
each other. Gamma spectroscopy systems are designed and adjusted to
produce symmetrical peaks of the best possible resolution. The peak
shape is usually a Gaussian distribution.
In most spectra the horizontal position of the peak is determined by
the gamma ray's energy, and the area of the peak is determined by the
intensity of the gamma ray and the efficiency of the detector.
The most common figure used to express detector resolution is full width at half maximum
(FWHM). This is the width of the gamma ray peak at half of the highest
point on the peak distribution. Resolution figures are given with
reference to specified gamma ray energies. Resolution can be expressed
in absolute (i.e., eV
or MeV) or relative terms. For example, a sodium iodide (NaI) detector
may have a FWHM of 9.15 keV at 122 keV, and 82.75 keV at 662 keV. These
resolution values are expressed in absolute terms. To express the
resolution in relative terms, the FWHM in eV or MeV is divided by the
energy of the gamma ray and multiplied by 100. Using the preceding
example, the resolution of the detector is 7.5% at 122 keV, and 12.5% at
662 keV. A germanium detector may give resolution of 560 eV at 122 keV,
yielding a relative resolution of 0.46%.
Detector efficiency
Not all gamma rays emitted by the source that pass through the
detector will produce a count in the system. The probability that an
emitted gamma ray will interact with the detector and produce a count is
the efficiency of the detector. High-efficiency detectors
produce spectra in less time than low-efficiency detectors. In general,
larger detectors have higher efficiency than smaller detectors, although
the shielding properties of the detector material are also important
factors. Detector efficiency is measured by comparing a spectrum from a
source of known activity to the count rates in each peak to the count
rates expected from the known intensities of each gamma ray.
Efficiency, like resolution, can be expressed in absolute or relative
terms. The same units are used (i.e., percentages); therefore, the
spectroscopist must take care to determine which kind of efficiency is
being given for the detector. Absolute efficiency values represent the
probability that a gamma ray of a specified energy passing through the
detector will interact and be detected. Relative efficiency values are
often used for germanium detectors, and compare the efficiency of the
detector at 1332 keV to that of a 3 in × 3 in NaI detector (i.e., 1.2×10
−3 cps/Bq
at 25 cm). Relative efficiency values greater than one hundred percent
can therefore be encountered when working with very large germanium
detectors.
The energy of the gamma rays being detected is an important factor in
the efficiency of the detector. An efficiency curve can be obtained by
plotting the efficiency at various energies. This curve can then be used
to determine the efficiency of the detector at energies different from
those used to obtain the curve. High-purity germanium (HPGe) detectors
typically have higher sensitivity.
Tidak ada komentar:
Posting Komentar